EN-14
Defender 6000 Indicators
Load Cell
Terminal
Blocks
Main Board Wiring Connections
Once the i-DT61PW and i-DT61XWE enclosure is opened, connections can be made to the terminal blocks on
the main board as shown in Figure 2-2.
i-DT61PW
i-DT61XWE
Figure 2-2
Jumper Connections
The i-DT61PW and i-DT61XWE indicators are designed to support both 2mV/V and 3mV/V load cells from the
same circuitry. A load cell output rating selection jumper is not required.
Figure 2-3 shows the terminal definitions for the analog load cell terminal blocks. Note that when using four
-wire load cells, jumpers must be placed between the +Excitation and +Sense terminals and between the
-Excitation and -Sense terminals.
Figure 2-3 Jumper Connections
After wiring is completed, replace the indicator housing screws. Make sure the water-proof cable gland is
properly tightened.
2.3.3 Communication Interface Cable to i-DT61PW
Attach the IR Communication cable (P/N: 30572910) to the indicator front panel, make sure the two holes in the
interface cable module match the two bolts which located in the front panel.
Pin
Connection
J3-1
+EXC
J3-2
+SEN
J3-3
+SIN
J3-4
GND
J3-5
-SIN
J3-6
-SEN
J3-7
-EXC
Figure 2-4
J3
-1
J3
-2
J3
-3
J3
-4
J3
-5
J3
-6
J3
-7