Defender 6000 Indicators
EN-33
4.4 Readout Menu
Enter this menu to set user preferences. Default settings are in
bold
.
Menu
Sub-Menu
Sub-Menu
(in
segment)
Options
Options ( in
segment)
Read Out
(
r.E.A.d
)
Reset
reset
no
, yes
NO
,
yes
Stability
Stable
0.5d,
1d
, 2d, 5d
0.5d
,
1d
,
2d
,
5d
Zero Range
ZErO
2%,
100%
2
,
100
Filter Level
FIltEr
Low,
Medium
, High
LOwW mMEd
,
HIGH
Auto Zero Track
AZt
Off,
0.5d
, 1d, 3d
OFF
,
0.5d
,
1d
,
3d
Bright Level (i-DT61XWE)
LIGHt
Low,
Medium
, High
mMEd
,
HIGH
,
LOwW
Back Light (i-DT61PW)
B.LIGHt
Off
, On, Auto
OFF
,
ON
,
AUtO
Screen
S.SAuEr
Off, 1min, 2min,
5min
(i-DT61PW)
Off, 5min, 10min,
30min
(i-DT61XW)
OFF
,
1
,
2
,
5
(i-
DT61PW)
OFF
,
5
,
10
,
30
(i-
DT61XW)
Auto Off
A.Off
Off
, 5min, 10min, 30min
OFF
,
5
,
10
,
30
P.SAVE (i-DT61PW)
P.SAuER
ON
, OFF
ON
,
OFF
L.KEY (i-DT61XWE)
L.KEY
Off
, 0.5min, 1min, 2min,
5min
OFF
,
0.5
,
1
,
2
,
5
End
End
\
Reset
[
reset
]
Reset the readout menu to factory defaults.
NO
= do not reset.
YES
= reset
Stability
[
StAblE
]
Set the amount the reading can vary before the stability symbol turns off.
0.5d
= 0.5 scale division
1d
= 1 scale division
2d
= 2 scale division
5d
= 5 scale division
Zero
[
ZErO
]
Set the percentage of scale capacity that may be zeroed.
2%
= zero range is +/-2%
100%
= zero range is +/-100%
Filter
[
FILtEr
]
Set the amount of signal filtering.
LOwW
= faster stabilization time with less stability.
mMEd
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
AZT
[
AZt
]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been exceeded.
Summary of Contents for Defender i-DT61PW