EN-22 Defender 5000 Indicators
3.4.3 Capacity
Set the scale capacity from 1 to 999999.
>|1|< Capacity
Specify the weight capacity for interval 1. If
Single
interval is enabled, this will be the scale capacity.
If
Dual
interval is enabled, this will be the first range.
>|2|< Capacity
Specify the weight capacity for interval 2. If
Dual
interval is enabled, this will be the scale capacity
and it must be bigger than >|1|< Capacity. If
Single
interval is enabled, this parameter will not be
shown.
3.4.4 Graduation
Set the scale readability from 0.0001 to 100.
>|1|<Graduation
Specify the graduation for weighing interval 1. If
Single
interval is enabled, this will be the graduation
for the entire weighing range of the scale. If Dual interval is enabled, this will be the graduation used
in the lower interval.
>|2|<Graduation
Specify the graduation for interval 2. If
Dual
interval is enabled, this will be the graduation for the
second weighing range of the scale. If
Single
interval is enabled, this parameter will not be shown.
NOTE:
Graduation settings are limited to values from Capacity divided by 600 to Capacity divided
by 75000. Therefore, not all settings are available for each capacity.
3.4.5 Language
Set the language displayed for menus and displayed messages.
English
Deutsch
Français
Italiano
Polski
Spanish
한국
中文
日本语
3.4.6 Power On Zero
Zero the balance at Power On.
Off = disabled.
On
= enabled.
3.4.7 Power On Unit
Set the unit that will be displayed at Power On.
Automatic
g
kg
lb
oz
lb:oz
t (Metric Tonne)
ton (Short Ton)