Defender 5000 Indicators
EN-17
3.2 Menu Navigation
To enter the Main Menu, press the button
from any application home screen.
Changing Settings
To change a menu setting, navigate to that setting using the following steps:
1. Enter the Menu
From any Application screen, press the
button. The Main Menu List appears on the screen.
2. Select the Sub-menu
Scroll to the desired Sub-menu in the Main Menu List using the Softkey corresponding to the icon
. Press the Softkey corresponding to the icon
to display the Sub-menu items.
3. Select the Sub-Menu Item
Scroll to the desired Sub-menu Item using the Softkey corresponding to the icon
.
Press the Softkey corresponding to the icon
to view the Sub-
menu item’s settings.
4. Select the Setting
Scroll to the desired Setting using the Softkey corresponding to the icon
.
Press the Softkey corresponding to the icon
to select the setting.
Press the Softkey corresponding to the icon
to return to the previous screen.
Press the Softkey corresponding to the icon
to exit the menu and return to the last active
Application mode.
3.3 Calibration Menu
The TD52 indicator offers three calibration methods:
Zero Calibration, Span Calibration and Linearity Calibration.
NOTES
:
1. Make sure that appropriate calibration masses are available before calibration.
2. Make sure that the scale base is level and stable during the entire calibration process.
3. Calibration is unavailable with LFT set to ON.
4. Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature.
5. To abort calibration, press the Softkey corresponding to the icon 'X'
anytime during the calibration
process.
6. When any selection within the GMP menu is enabled, calibration results are automatically printed.
3.3.1 Zero Calibration
Zero calibration uses one calibration point. The zero calibration point is established with no weight
on the scale. Use this calibration method to adjust for a different dead load without affecting the span
or linearity calibration.
Calibration procedures:
Long press the button
to enter the Main
Menu. Press the Softkey corresponding to the
icon
to enter the Calibration sub-menu.