![OHAUS Adventurer Pro Grain Instruction Manual Download Page 7](http://html.mh-extra.com/html/ohaus/adventurer-pro-grain/adventurer-pro-grain_instruction-manual_726739007.webp)
Adventurer Pro Grain Test Scale
EN-5
The scale performs a segment check. The scale then
displays the last selected application mode.
To turn power off, press and hold the
Off
button until OFF
is displayed, then release the button.
2.5.2 Turning Power On and Off
Press the
On O/T
button to turn power on.
2.6 Initial Calibration
When the scale is first installed, and when it is moved to another location, it must be calibrated
to ensure accurate weighing results. Have the appropriate calibration masses available before
beginning calibration. Refer to Section 3.7.1and 3.7.2 for masses and calibration procedure.
2.3 Selecting the Location
Locate the scale on a firm, steady surface. Avoid locations with excessive air current, vibrations,
heat sources, or rapid temperature changes.
2.4 Leveling the Scale
Before the scale is used, the feet should be adjusted so that the scale is level with the surface.
This will enable accurate weighing. You will see a level bubble indicator in a small round
window on the front of the scale. Level the scale by adjusting the leveling feet so the bubble is
centered in the circle, as seen below.
Balance Level
2.5 Connecting Power
2.5.1 AC Adapter
Connect the AC Adapter to the wall outlet. Connect the plug into the receptacle on the rear of
the scale.
For use with CSA Certified (or equivalent approved) power source, which must
have a limited circuit output.
3 page 3 to 6.pmd
7/27/2006, 3:11 PM
5