![OHAUS Adventurer Pro Grain Instruction Manual Download Page 20](http://html.mh-extra.com/html/ohaus/adventurer-pro-grain/adventurer-pro-grain_instruction-manual_726739020.webp)
Adventurer Pro Grain Test Scale
EN-18
First the zero reading is taken. Next the display shows the
first calibration weight value. PUT WEIGHT (blinking) is
displayed. Place the specified calibration weight on the
pan.
After a few seconds, place the second specified calibration
weight on the pan. PUT WEIGHT (blinking) is displayed.
After a few seconds, the calibration weight is displayed.
LINEAR DONE is momentarily displayed when calibration is
completed.
After calibration, remove weight from pan. To leave the
menu, press
Exit
.
To abort calibration at any time, press the
Exit
button.
3.7.2 Calibration Masses
Model
Linearity
Span
Weight Class
Calibration Points
Calibration Points
ASTM OIML
AVD4102N
2000g / 4000g
2000g, 3000g,
4000g
1
E2
Calibration Test
Calibration test allows a check of a known calibration mass
against the last stored calibration information in the scale.
Press and hold the
Menu-Cal
button until MENU CALIBRATE
is displayed.
Press
Yes
to enter calibration.
Press
No
until CAL TEST is displayed.
Press
Yes
and follow the screen instructions.
Place the specified weight on the pan. After a short period,
the display indicates the difference in weight from the last
calibration, and then will display the calibration weight on
the pan. After the test, remove the calibration weight from the
pan. To leave the menu, press
Exit
.
Linearity Calibration (Cont.)
6 page 16 to 18.pmd
7/27/2006, 3:11 PM
18