Adventurer Pro Grain Test Scale
EN-17
With the scale on, press and hold the
Menu-Cal
button
until MENU is displayed, then release the button. CALIBRATE
(blinking) is displayed.
Press the
Yes
button to enter span calibration.
Press the
Yes
button to initiate span calibration.
First the zero reading is taken. Next the display shows the
default span calibration weight value.
Press
No
to change to an alternate calibration weight. Press
Yes
when the desired calibration weight is displayed.
Place the specified calibration weight on the pan. After
calibration, remove the weight from the pan.
To abort calibration at any time, press the
Exit
button.
Span Calibration
NOTICE:
Linearity Calibration is required only if the linearity error exceeds tolerance in
specification table.
Linearity Calibration
With the scale on, press and hold the
Menu-Cal
button
until MENU CALIBRATE is displayed, then release the button.
CALIBRATE (blinking).
Press the
Yes
button to enter calibration.
Press the
Yes
button and follow the screen instructions.
Use the weight values shown on the primary display.