Defining Custom Unit
Set Custom to ON in the Unit menu to enable
and define the Custom Unit. The custom unit is
defined by a conversion factor and a least
Significant Digit (LSD). The conversion factor
is used by the balance to convert grams to the
custom weighing unit and is defined by entering
a factor and an exponent. The factor is a value
between 0.1000000 and 1.999999 inclusive.
The exponent moves the decimal point of the
factor to the right for positive values or to the left
for negative values.
Custom Unit = Conversion Factor x Grams.
The LSD is the value by which the displayed
weight is incremented or decremented.
LSD
Result
.5
Adds one decimal place
Display counts by 5
1
Display counts by 1
2
Display counts by 2
5
Display counts by 5
10
Display counts by 10
100
Display counts by 100
End Unit
Press the
Yes
button to advance to the Print-1
submenu. Press
No
to return to the first available
unit.
3.7.7
Print-1 and Print-2
The Print-1 submenu is used to set printing
parameters for an external printer or computer.
Print-2 contains an identical menu and is used
when the optional COM2 port is installed.
Enter the Factor, following the instructions for
Editing Limits in Section 3.5.4. Enter the
Exponent and LSD using the
Yes No
buttons.
Factor
Exponent
Conversion
(+3 to -3)
Factor
.1234
3
123.4
.1234
2
12.34
.1234
1
1.234
.1234
0
.1234
.1234
-1
.01234
.1234
-2
.001234
.1234
-3
.0001234
PRINT-1
Output
When Stable
On/
Off
GLP Tare
On/
Off
Auto Print
Off
, Cont., Interval, When Stable
1 -3600 seconds
Stable- Load, Load & Zero
Content Num Only-
Off
, Header, Weight,
Ref-
Off
, Result-
On
,GLP-
Off-
all others
Off
Layout
Line Format -
Multi
, Single
4LF-Yes/
No
Form Feed-Yes/
No
List
Yes/
No
End Print-1
Output
Set When Stable to On to print only stable values.
Set When Stable to Off to print stable or unstable
values.
Set GLP Tare to On to print GLP data once after
a tare operation. Set GLP Tare to Off to disable
this feature.
Auto Print
When set to Continuous, the displayed value is
printed continuously. When set to Interval, the
displayed value is printed at the user specified
time interval (1 to 3600 seconds).
If set to When Stable, the scale will
automatically print the displayed value when
stability is achieved. An additional setting
must be made to determine if only stable non-
zero values will be printed (load setting) or if
stable zero and non-zero values will be printed
(Load & Zero). When set to Off, the Auto print
feature is disabled.