7000 SERIES INDICATORS
EN 113
Index
Name
Description
Notes
032
Output 1
Control
0
– Turn DIO Output 1 OFF
1
– Turn DIO Output 1 ON
033
Output 2
Control
0
– Turn DIO Output 2 OFF
1
– Turn DIO Output 2 ON
034
Output 3
Control
0
– Turn DIO Output 3 OFF
1
– Turn DIO Output 3 ON
035
Output 4
Control
0
– Turn DIO Output 4 OFF
1
– Turn DIO Output 4 ON
037
Restart
Terminal
1
– Restart Terminal
040
Zero
Adjustment
1
– Trigger Zero Adjustment
041
Mid span
adjustment
1
– Trigger mid span adjustment
Used when linearity
is enabled as the
middle span point
042
Span
adjustment
1
– Trigger high span adjustment
049
Ethernet test
1
– enter Ethernet test
0
– exit Ethernet test
050
USB test
1
– enter USB test
0
– exit USB test
051
Mode switch
1
– Switch to Normal Mode
0
– Switch to Application Mode
Setup
– Scale (read and write)
Index
Name
Description
100
Scale Block
Entire Scale Block
101
Scale Name
20 characters max. No entry on
terminal.
102
Approval Type
0
– None
1
– Argentina
2
– Australia
3
– Canada
4
– OIML
5
– USA
103
Certificate #
Manual entry - 20 characters max.
104
Unit
1
– g
2
– kg
3
– lb
4
– t
5
– ton
105
Number of ranges
0
– One range
1
– 2 ranges
106
Range 1 capacity
Manual entry