7000 SERIES INDICATORS
EN 107
Response:
T_I
Command understood, not executable at present.
Response:
T_+
Upper limit of zero setting range exceeded.
Response:
T_-
Lower limit of zero setting range exceeded.
Comments
The existing tare will be overwritten and replaced by the new preset tare weight value.
The terminal will wait for up to 3 seconds after receiving a “T” command for no-motion. If motion does not settle
within this time, the command is aborted and the T_I response is sent.
TA
– INQUIRE/ENTER TARE VALUE
Command:
TA
Inquiry of tare weight value
TA_Tare Preset Value_Unit
Entry of a preset tare value.
Response:
TA_A_TareWeightValue_Unit
Current Tare weight value.
Response:
TA_I
Current Tare weight value cannot be transferred (the terminal is currently executing
another command, such as zero setting).
Response:
TA_L
Command understood, invalid parameter.
Example
Command:
TA_10.00_kg
Load a preset tare of 10 kg.
Response:
TA_A_ _ _ _ _10.00_kg
The 10.00 kg tare value was accepted.
Comments
The existing tare will be overwritten and replaced by the new preset tare weight value.
The terminal will automatically round the tare value to the current readability.
The preset value must be entered in primary units.
TAC
– CLEAR TARE VALUE
Command:
TAC
Response:
TAC_A Tare value cleared.
Response:
TAC_I Command understood, not executable at present.
TI
– TARE IMMEDIATELY
Command:
TI
Response:
TI_S_WeightValue_Unit Taring performed, stable tare value.
Response:
TI_D_WeightValue_Unit Taring performed, non-stable (dynamic) tare value.
Response:
TI_I
Command understood, not executable at present.
Response:
TI_L
The command is not executable.
Response:
TI_+
Upper limit of taring range exceeded.