7000 SERIES INDICATORS
EN 105
S_S_ _ _ _ _ 129.09_kg
S_S_ _ _ _ _ 129.09_kg
S_D_ _ _ _ _ 114.87_kg
. . . The scale continues to send stable or dynamic weight values
Comments
The SIR command is overwritten and cancelled by the commands S, SI, SR, and @.
The data output rate is approximately 4 per second.
Z
– ZERO
Command:
Z
Response:
Z_A
Command performed, meaning the scale was in gross mode, scale was stable and weight
was within the zero capture range.
Response:
Z_I
Command understood, not executable at present.
Response:
Z_+
Upper limit of zero setting range exceeded.
Response:
Z_-
Lower limit of zero setting range exceeded.
Comments
The calibrated zero point determined during calibration is not influenced by this command.
The terminal will wait for up to 3 seconds after receiving an “Z” command for no-motion. If motion does not settle
within this time, the command is aborted and the Z_I response is sent.
@
– RESET
Command:
@
Response:
I4_A_“12345678”
12345678
Serial number of the scale
Comments
Resets the scale to the condition found after switching on, but without a zero setting being performed.
All commands awaiting responses are canceled.
The tare register is cleared.
The SIR and SR commands are cancelled.
The Reset command is always executed except that a Reset command received by the terminal during the
calibration and test procedure cannot be processed.