11.1.23 Vit-PN module
VX210157
Quantity
Description
1
Vit-PN module
4
M3x10 oval head screw, TORX, ecosyn-fix
STEP 1:
1.
2.
Remove system PCB according to
Remove board carrier by unfas-
tening the 4 screws.
STEP 2:
3.
4.
5.
Disconnect the 2 tubes from the air
module.
Disconnect the green and black
tubes from Vit-PN outputs.
Disconnect the black tube from the
Visco module.
STEP 3:
6.
Unfasten the 4 screws from carrier
by using a tool with a long shaft.
Summary of Contents for OS 4