2021-03-23 / Rev. 10
VV016042
99
11.3.5
Endo laser module
VX210162
Quantity
Description
1
Endo laser module
STEP 1:
1.
2.
Remove housing according to sec-
tion
11.1.5 step 3.
Remove isolation transformer as-
sembly to get access to the laser
module.
Remove the top screw completely
(TX30). The bottom screw needs
only to be loosened. Swing the
transformer to the left side.
STEP 2:
3.
Remove 2 (TX10) screws from the
power supply carrier.
STEP 3:
4.
5.
Remove 8 (TX25) screws from both
supports where one of them holds
the supply PCB in place.
Push the supports out of the frame.