2021-03-23 / Rev. 10
VV016042
37
9.7.2
Fibre check
If the deviation is higher than expected, check power directly at the laser module:
STEP 1:
1.
2.
3.
4.
5.
6.
7.
Disconnect fibre cable from laser
module.
Protect end of fibre cable with the
cap attached to the fibre itself.
Connect test fibre of measuring sen-
sor head to laser module.
Set external fibre detection to
“NO”
(see
Measure power and compare it with
the intended value.
If the value is within the tolerance,
exchange the fibre (see
Calibrate as described in
If the power at the module itself matches the values submitted at the software, then the power loss is caused by the fibre
(narrow radii or kinked).
Summary of Contents for OS 4
Page 1: ......
Page 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Page 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Page 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Page 151: ...VV016042 2021 03 23 Rev 10 ...