Fault
Possible cause
Remedy
Reference
section *
Oil in intake line
Oil comes from the vacuum system.
Check the vacuum system.
or in vacuum
Anti-suckback valve is obstructed.
Clean or repair the valve.
5.4-G
vessel.
Sealing surfaces of anti-suckback valve
Clean or repair the intake port and valve.
5.4-G
are damaged or dirty.
Oil level is too high.
Drain the excess oil.
5.4-B
Pump’s oil
Exhaust filters are clogged or damaged.
Replace the filters.
5.4-C
consumption too
Nozzle of float valve is clogged.
Check the valve, clean the nozzle.
5.4-I
high, oil mist at
Oil level is too high.
Drain the excess oil.
5.4-B
exhaust.
Oil is turbid.
Condensation.
Degas the oil or change the oil and clean
4.1/5.4-B
the pump.
Precaution: open the gas ballast valve or
insert a condensate trap.
Clean the gas ballast intake filter.
5.4-G
Pump is
Oil level is very low
Add oil.
5.4-B
excessively noisy.
(oil is no longer visible).
Oil filter is clogged.
Change the oil and filter.
5.4-B
Coupling elements worn
Install new coupling element
5.4-J
Large vacuum leak in system.
Repair vacuum leak.
5.4-J/K
* Reference section: This coluum refers to the section in the Operating Instructions that contains the applicable repair information.
1) Bubble test : Let the pump run at operating temperature with degassed oil, without gas ballast and with closed intake port.
Immerse the exhaust line in a bucket of water. If a steady stream of bubbles appear, the pump has an external leak.
NB: For any other operation as described before, return the pump to a Œrlikon Leybold Vacuum service center.
Never mount used seals. Always mount new seals.
Summary of Contents for 10926
Page 36: ...36 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 6 SV200 SV300 similar Spare parts ...
Page 38: ...38 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum Spare parts fig 7 SV200 SV300 similar ...
Page 40: ...40 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 7 SV200 SV300 similar Spare parts ...
Page 42: ...42 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum ...