27
GAET02307_002_10 - 06/2008 - © Œrlikon Leybold Vacuum
Maintenance
5.3 Œrlikon Leybold Vacuum Service
Whenever you send us in equipment, indicate whether the equipment is
contaminated or is free of substances which could pose a health hazard. If
it is contaminated, specify exactly which substances are involved. You must
use the form we have prepared for this purpose.
A copy of the form has been reproduced at the end of these Operating
Instructions: “Declaration of Contamination for Compressors, Vacuum
Pumps and Components“. Another suitable form is available from
www.oerlikon.com
Œrlikon Leybold Vacuum Systems
Documentation
Download Documents.
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal
requirements and for the protection of our employees.
We must return to the sender any equipment which is not accompanied by a
contamination statement.
The pump must be packaged in such a way that it will not be damaged
during shipping, and so that no harmful substances can escape from the
package.
When disposing of used oil, please observe the relevant environmental
regulations.
5.4 Maintenance Work
Checking the oil
A. Oil level
The pump oil level during operation must always be between the middle and
top edge of the oil-level glass (6/4).
When necessary, switch off the pump and add the correct quantity of oil.
Overfilling leads to oil losses at high intake pressures.
High oil consumption often indicates that exhaust filters are clogged.
Stop the pump for filling oil.
The oil level should be checked at least once a day.
B. Oil Condition GS77
Normally the oil is clear and transparent. If the oil darkens, it should be
changed.
If gases or liquids dissolved in the oil result in deterioration of the ultimate
pressure, oil can be degassed by allowing the pump to run for about 30 min.
with the intake port closed and the gas ballast valve open.
The amount of oil required for an oil check should be drained via the oil-drain
lug (6/5) into a beaker or similar container with the pump switched off but still
at operating temperature.
Observe the safety regulations!
Caution
Warning
Formulaire
Contamination
Summary of Contents for 10926
Page 36: ...36 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 6 SV200 SV300 similar Spare parts ...
Page 38: ...38 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum Spare parts fig 7 SV200 SV300 similar ...
Page 40: ...40 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 7 SV200 SV300 similar Spare parts ...
Page 42: ...42 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum ...