22
GAET02307_002_10 - 06/2008 - © Œrlikon Leybold Vacuum
Installation
Caution
3.3 Start-up
The pumps are supplied with the necessary oil filling in ready-to-use
condition. Always verify proper oil level before operating the pump.
The normal oil level is in the middle of sight glass.
If oil has to be added, unscrew the oil-fill plug, add oil and screw the plug
firmly back in.
The Sogevac
®
pump is designed for normal startup at temperatures over
12°C (54°F) (as per PNEUROP). With adequate oils, cold starts are possible
above 0°C (32°F).
■
Before starting the pump ensure that the attached accessories meet the
requirements of your application and that safe operation is ensured.
■
Never expose part of the body to the vacuum. There is a danger of injury.
Never operate the pump with an open and thus accessible inlet. Vacuum
connections as well as oil filling and oil draining openings must not be open
during operation of the pump.
■
The safety regulations which apply to the specific application in each case
must be observed. This applies in particular to installation, operation and
maintenance (servicing) as well as waste disposal and transportation.
Do not lay the hand on the intake to check succion.
Exposure of a part of the body to the vacuum result
s
in a rush of blood in the
exposed part.
Warning
Summary of Contents for 10926
Page 36: ...36 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 6 SV200 SV300 similar Spare parts ...
Page 38: ...38 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum Spare parts fig 7 SV200 SV300 similar ...
Page 40: ...40 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum fig 7 SV200 SV300 similar Spare parts ...
Page 42: ...42 GAET02307_002_10 06 2008 Œrlikon Leybold Vacuum ...