background image

(

)

3-4. Programming Over Voltage Protection(OVP)

21

3-5. Programming Over Current Protection(OCP)

23

3-6. Programming Under Voltage Limit(UVL)

25

3-7. Programming Over Voltage Limit(OVL)

26

3-8. Programming Under Current Limit(UCL)

27

3-9. Programming Over Current Limit(OCL)

28

29

-------------------

-------------------

-------------------

-------------------

-------------------

-------------------

3 10 KEY LOCK

29

30

GPIB Set-Up
GPIB Installation Set-Up
RS232C Set-Up
RS232C Installation Set-Up

3-12. V/I & LMT DISPLAY

34

---------------------------------------------

----------------------------------------------------

----------------------------------------------------

3-10. KEY LOCK
3-11. IO/Local

LMT DISPLAY Function
V/I Function

3-13. Output ON/OFF

35

3-14. Error Message Display

35

ERROR Check

---------------------------------

-----------------------------------------------

4. CALIBRATION

36

4-1. Characteristics

36

4-2. Preparing Accurate Calibration

36

4-3. Performance of Measuring Instrument for Accurate Calibration

--- 37

4-4. Measuring Technique

37

Equipment Connection

--------------------------------------------------

-----------------------------

----------------------------------

----------------------------------------------------

Electronic Load

Current-Monitoring Resistor(shunt)

DVM(Digital Volt Meter)
Programming
Measuring Position

4-5. Calibration Using Front Panel

39

CALIBRATION KEY Structure

----------------------------

CALIBRATION KEY Structure
Voltage CALIBRATION Operation
Current CALIBRATION Operation

4-6. Calibration Using REMOTE INTERFACE

43

Connecting Measuring Instrument
Remote Calibration Command Sequence
Voltage CALIBRATION Operation
Current CALIBRATION Operation

-------------------

 -  3  - 

Summary of Contents for OPX-93D

Page 1: ...Programmable DC Power Supply User Manual www odacore com OPX 93D Manual Part NO 018OPX 93D 2 0...

Page 2: ...tained herein Should ODA and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agr...

Page 3: ...a hazard that could result in personal injury CAUTION notice until the indicated conditions Verify that all safety precautions are taken CAUTION Before Applying Power are fully understood and met saf...

Page 4: ...letely disconnected from ac mains when switch is off In position of a bi stable push switch Direct current Earth ground terminal Frame or chassis terminal Protective earth ground terminal Alternating...

Page 5: ...fore Input Power 8 Check Power Cord Check AC Input Power Switch 1 6 Check After Input Power 9 Check Sequence Earliest Setting Value When Last state Mode is Disabled 1 7 Product Installation 10 Cooling...

Page 6: ...OFF 35 3 14 Error Message Display 35 ERROR Check 4 CALIBRATION 36 4 1 Characteristics 36 4 2 Preparing Accurate Calibration 36 4 3 Performance of Measuring Instrument for Accurate Calibration 37 4 4 M...

Page 7: ...Operation 5 10 CAL RESTORE 53 CAL RESTORE Operation 5 11 CAL BACKUP 53 5 11 CAL BACKUP 53 CAL BACKUP Operation 5 12 CAL DEFAULT 54 CAL DEFAULT Operation 5 13 LOAD DEFAULT 54 LOAD DEFAULT Operation 6...

Page 8: ...68 7 1 Operational Error 68 7 2 Hardware Error 68 68 7 4 Calibration Error 70 7 5 Fixed Memory Check Error 70 7 3 Remote Calibration Error 7 6 Interface Commands Error 71 8 Specifications 72 9 Warning...

Page 9: ...OVP OCP Store Recall up to 10 Error Messages 1 2 19inch 3U Rack Compact Size Remote Interface Characteristics RS232C GPIB Interface Support SCPI Standard Commands for Programmable Instruments Compatib...

Page 10: ...REAR Terminal is damaged Check Scratches on the product body Check Scratches and LCD Dot shatter Check Scratches and LCD Dot shatter Biographic Check Check Alarms when the Power is ON If there is an...

Page 11: ...ducts GND before using the goods Check AC Input Power Switch The product is designed single phase with AC100V AC220V 10 50 60H We set The product is designed single phase with AC100V AC220V 10 50 60Hz...

Page 12: ...or error code will be saved You could press ERR Key to see the error code You could press ERR Key to see the error code Read Reference 7 Error Messages for further details Earliest Setting Value When...

Page 13: ...uce output current 0 70 When it is installed in Rack please consider about ventilation Diagram 1 3 Bottom view Bench Operation There should be spare spaces on products side front rear for ventilation...

Page 14: ...Lamp Channel1 P1 16Char Type LCD Display Channel2 P2 16Char Type LCD Display CV Mode Lamp CC Mode Lamp Left Displace Cursor Menu change Key OVP Set up Condition Lamp OCP Set up Condition Lamp Remote I...

Page 15: ...Lamp When in Over Current Protection is Set up the lamp will be lighted They can be changed in Factory Mode 7 Remote Interface Lamp When remote interface is recalled Lamp will be lighted Front panel c...

Page 16: ...value will fall If it rotate right CW value will rise in numerical menu mode or Voltage Current 16 Positive Output Output Terminal Voltage Current Positive 17 Negative Output Output Terminal Voltage...

Page 17: ...Left cursor Key increases the digit right cursor Key decreases the digit 5 To raise Limit value rotate encoder clock wise To reduce it rotate encoder anti clock wise 6 Check Set up value is successfu...

Page 18: ...ial number 2 Channel 1 P1 RS232C Port RS232C Interface Port which controls Channel 1 P1 3 Channel 1 P1 GPIB Port GPIB Interface Port which controls Channel 1 P1 4 Ch l 2 P2 RS232C P t 4 Channel 2 P2 R...

Page 19: ...Programming Under Voltage Limit UVL 6 Programming Under Voltage Limit UVL Explanation of programming under voltage limit 7 Programming Over Voltage Limit OVL Explanation of programming over voltage li...

Page 20: ...I LMT Key and move cursor to Current Use Cursor key and Encoder switch to set up higher current value than the load In Protection OCP UCL OCL mode current cannot rise more than Protection level if use...

Page 21: ...aise more than Protection Limit PROTECTION LIMIT M ill b di l d d t i P t ti L l l Message will be displayed and set up in Protection Level value To set up current use cursor key to change the current...

Page 22: ...pply s total output value Therefore if users try to print maximum voltage we cannot guarantee V sensing performance Power supply will be in Unregulated condition Al if h l d l d li i th 1V l b U l t d...

Page 23: ...sensing lead line safely to the load Power supply s programming feedback is a part of root might cause problems by using V Sensing DO NOT connect load lead line to sensing TERMINAL Connecting Remote...

Page 24: ...escription Changed Returns after displaying above Message To cancel To cancel please press I O LOCAL Key or wait for 10 seconds Wait 10 Sec LCD Display Description Cancel Note Read reference about ear...

Page 25: ...isable 5 OVP USE When OVP Trip occurs OVP TRIPPED message will be displayed To Clear OVP Trip firstly raise OVP Level To set up Level press PROTECT Key LCD Display Description PROTECT 1 OVP Press PROT...

Page 26: ...LCD Display Description Changed It returns when above message is displayed To cancel To cancel press I O Local Key or wait for 10 seconds wait 10sec LCD Display Description Cancel wait 10sec LCD Disp...

Page 27: ...TRIPPED message will be displayed To clear OCP trip firstly raise OCP Level value Press PROTECT Key to set up Level value LCD Display Description PROTECT 1 OVP Use right cursor key to move on to OCP M...

Page 28: ...t tries to rise more than protection level PROTECTION LIMIT will be displayed and voltage will be set up protection level value Press PROTECT Key to set up UVL Level LCD Display Description PROTECT 1...

Page 29: ...ies to rise more than protection level PROTECTION LIMIT will be displayed and voltage will be set up protection level value Press PROTECT Key to set up OVL Level LCD Display Description PROTECT 1 OVP...

Page 30: ...level value OVL If it tries to rise more than protection level PROTECTION LIMIT will be displayed and voltage will be set up protection level value Press PROTECT Key to set up UCL Level LCD Display De...

Page 31: ...OVL If it tries to rise more than protection level PROTECTION LIMIT will be displayed and current will be set up protection level value Press PROTECT Key to set up OCL Level LCD Display Description PR...

Page 32: ...ther keys in Front panel KEY LOCK will be displayed and keys will be locked To allow keys in Front panel press KEY LOCK Key KEY UnLOCK will be displayed and keys will be unlocked KEY UnLOCK will be di...

Page 33: ...it is not working even RMT lamp is off there may be possibility that it is in KEY LOCK system Please Unlock the KEY LOCK System GPIB Set Up Users are able to control 15 products including PC and commu...

Page 34: ...line should be shield line and should not be longer than 2M Also distance between the equipments should not longer than 20M total DO NOT connect more than 15 products including PC IEEE488 2 Connector...

Page 35: ...O LOCAL Key LCD Display Description BR 1 9600 bps can be different by condition Baud rate are 9600 19200 38400 57600bps they should be the same with Remote Interface Baud rate For example to select 38...

Page 36: ...Female type standard cross table wire diagram Read reference 1 2 Accessories Product Option Option to select wire distances Diagram 3 6 Diagram 3 6 It is convenient when user use adopt cable separatel...

Page 37: ...ge for 4 seconds return to previous page 4 seconds LCD Display Description OUTPUT OFF V I Function V I Function Move cursor to voltage or current Power Switch ON Check whether OUTPUT OFF is displayed...

Page 38: ...are stack structured and recent message will be accessed first They will be saved up to 10 from 11th message the earliest messages will be deleted When user check message by pressing ERR Key the messa...

Page 39: ...in fixed memory Support Backup Restore Calibration Data pp p Simple operation using 1 1 method of calibrating instrument Meter value 4 2 Preparing Accurate Calibration Please use measuring instrument...

Page 40: ...t On Off Resolution 10 uA Accuracy 0 01 Current monitoring ODA LP600 Current Calibration Power Supply Protection 0 01 0 01 While Current Calibration Voltage Monitoring Current Calibration Current moni...

Page 41: ...interface Current Monitoring Resistor shunt Use current monitoring resistor because general current measuring instrument is not adequate to high current capacity Select TCR less than 10ppm Use 0 01 h...

Page 42: ...th B measuring instrument must be connected to A Binding post neck Diagram 4 3 Front Panel Terminal Connections Side view 4 5 Calibration Using Front Panel Explanation about sequence of manual calibra...

Page 43: ...ption V LOW 0 014V Voltage value can be different Wait for voltage shown on DVM Meter to be settled Release Wait for 5 minutes Use cursor key and encoder switch to input voltage value to DVM Meter For...

Page 44: ...nect Power supply terminal to electronic load input terminal and connect power supply to input terminal current monitoring resistor and opposite side lead to power supply output terminal Connect DVM i...

Page 45: ...4 CAL CURR HIGH and press CAL Key LCD Display Description A HIGH 3 2778A Current can be different Wait for voltage shown on DVM Meter to be settled Wait for 5 minutes If it is stabilized calculate cu...

Page 46: ...from the first If an error occurs Remote Calibration should be done from the first Voltage CALIBRATION Operation Switch OFF the Electronic Load Send Voltage Minimum Calibration command to power supply...

Page 47: ...erefore please set up more than the maximum current Wait for 5 minutes and measure DVM current Send power supply s current value from prepared current measuring program For example If measured value i...

Page 48: ...ng time it can change voltage current into minimum value When the product is not used for a long time it can change voltage current into minimum value When the product is not used for a long time it c...

Page 49: ...ng Maximum Value OVP Level OVP Setting Maximum Value OCP Level OCP Setting Maximum Value UVL Level 0V OVL Level Limit Maximum Value UCL Level 0A OCL Level Limit Maximum Value Output Mode OFF USER MEM...

Page 50: ...tput OFF FULLY Restore previous value before it was OFF It restores Output setting also can be dangerous because when it is booted it is starting with output ON LAST STATE Operation Power Switch ON wh...

Page 51: ...UTO CURSOR MV Operation Power Switch ON while pressing FACTORY Key Holding Power Switch ON When 1 USER MEM CLEAR Message displayed release FACTORY Key Use cursor key to move on to 3 AUTO CURSOR MV Men...

Page 52: ...ABLE Auto front panel lock mode is enabled AUTO KEY LOCK Operation Power switch ON while pressing FACTORY Key Holding Power Switch ON When 1 USER MEM CLEAR Message is displayed release FACTORY Key Use...

Page 53: ...OVP Enabled OVP USE Operation Power Switch ON while pressing FACTORY Key Holding Power Switch ON When 1 USER MEM CLEAR Message is displayed release FACTORY Key Use cursor key to move on to 5 OVP USE M...

Page 54: ...BLE OCP Enabled OCP USE Operation Power Switch ON while pressing FACTORY Key Holding Power Switch ON When 1 USER MEM CLEAR Message is displayed release FACTORY Key Use cursor key to move on to 6 OCP U...

Page 55: ...n every 50ms 50Hz Measure voltage or current in every 20ms Default setting 100Hz Measure voltage or current in every 10ms 300Hz Measure voltage or current in every 3 3ms 1 3KHz Measure voltage or curr...

Page 56: ...D Display Description Done Release 5 11 CAL BACKUP To obtain guarantee of high accuracy the product should be calibrated every 6 month at institute Also to use without problem it should be calibrated...

Page 57: ...TORY Key Holding Use cursor key to move on to 10 CAL DEFAULT Menu Press FACTORY Key to restore LCD Display Description Done 5 13 LOAD DEFAULD Function which bring back setting value changed in Factory...

Page 58: ...gle bracket is option or parameters Parameters inside Braces cannot be jumped Triangle bracket are changeable 2 Bar or more indicates the more parameter Command letters are standard LF 0x0A Sending at...

Page 59: ...rrent MIN MAX Calibration Commands FACT USER M CLR FACT OVP DIS ENA FACT LAST STA DIS SAF FUL FACT OVP FACT LAST STA FACT OCP DIS ENA FACT AUTO CUR DIS ENA FACT OCP FACT AUTO CUR FACT ADC 5 20 50 100...

Page 60: ...ent using PC Remote Interface SOURce VOLTage voltage UP DOWN U bl t t t t lt d t d t lt d t ill b t Users are able to set output voltage and current product voltage and current will be set Before putt...

Page 61: ...STATe SOURce VOLTage PROTection STATe This command checks OVP function activated or deactivated Return value 0 OVP Function is deactivated 1 OVP Function is activated ex volt prot stat return value 1...

Page 62: ...p voltage first volt ovl 5 Set up voltage more than above voltage value From now on more than 5V cannot be set SOURce VOLTage OVL This command checks the OVL Over voltage limit Level This command chec...

Page 63: ...SOURce CURRent PROTection SOURce CURRent PROTection This command checks the OCP Over current protection Trip Level Return value numeric value ex curr prot return value 5 2000 SOURce CURRent PROTectio...

Page 64: ...UCL This command checks the UCL Under current limit Level Return value numeric value ex curr ucl return value 2 0000 SOURce CURRent OCL current This command set up OCL Over current limit Level numeric...

Page 65: ...quence CAL VOLT MIN Operate Low value CAL VOLT voltage Send measured voltage to DVM CAL VOLT MAX Low value is done operate High value CAL VOLT voltage Send measured voltage to DVM When voltage value i...

Page 66: ...command which checks Last state condition value Return value DISABLE SAFETY FULLY ex fact last sta return value SAFETY FACT AUTO CUR DIS ENA This command makes voltage current range to minimum when f...

Page 67: ...abled or disabled DIS Disable OCP function is disabled ENA Enable OCP function is enabled ex fact ocp dis Set up OCP Function to be disabled FACT OCP It is command which checks whether OCP is enabled...

Page 68: ...alarm once ex syst beep It will be alarmed SYSTem ERRor It is command which checks error messages from power supply Error messages will be saved up to 10 fixed memory from 11th error the lastest erro...

Page 69: ...rint 1 OR ON Output Allowed 0 OR OFF Output shut off 0 OR OFF Output shut off ex1 outp on Output Allowed ex2 outp off Output shut off OUTPut STATe It is command which check power supply output conditi...

Page 70: ...er 1 10 memories 1 10 Memory saving range ex sav 2 Save in number 2 memory RCL 1 2 3 4 5 6 7 8 10 It is command which applys saved User Memory to power supply Memories are selectable from 1 to 10 Memo...

Page 71: ...rt s circuit is failed 202 Front panel operating failed Front Panel does not reply 255 Error not define It occurs when a defined error occurs 7 3 R t C lib ti E 7 3 Remote Calibration Error Read refer...

Page 72: ...X VALUE 28 Ignored min save under curr It occurs when user sent commands about max value for current before sending minimum Sequence Min VALUE MAX VALUE 29 Invalid min value use under curr It occurs w...

Page 73: ...t checks data and inform you with error message 80 Memory limit volt error There is an error in voltage memory limit 81 Memory limit curr error There is an error in current memory limit 82 Memory max...

Page 74: ...ed command is sent ex volta 10 Recognize only volt or voltage Fix voltage 10 OR volt 10 220 No execution Occurs when a command cannot be operated ex Volt 20 Voltage setting 20V g g Volt OVP 15 Error o...

Page 75: ...8 Specifications Below is specifications of 19 inch 1U 600W 2 5KW product visit www odacore com to download 72...

Page 76: ...ritical damage DO NOT provide vibration or hard physical impact Th h ld b h d til t There should be enough spaces around ventilator DO NOT place heavy material object on the product Please aware devic...

Page 77: ...62 Bupyeong daero 329beon gil Bupyeong gu Incheon city 403 858 Korea TEL 82 2 1800 8644 FAX 82 32 715 5456 www odacore com sales2 odacore com ODA TECHNOLOGIES CO LTD...

Reviews: