NUVE NC 23B User Manual Download Page 6

 

1.

 

INTRODUCTION 

1.1.

 

USE AND FUNCTION 

 NC  23B  Bench  Top  Steam  Sterilizers  are  widely  used  for  sterilization  of  appliances  utilized  in 
general medical applications, dentistry, acupuncture and veterinary sciences. These equipment 
are  also  appropriate  for  the  sterilization  of  tools  and  instruments  that  are  directly  in  contact 
with blood or other body fluids (e.g. tools and instruments used at beauty clinics, tattoo centers, 
hair-dressers etc.) 

NC  23B  Steam  Sterilizers  are  manufactured  in  accordance  with  EN  13060  standard,  which 
outlines the  general  rules  for  small  steam  sterilizers  and determines  the  sterilization methods 
for  specific  sterilization  loads.  According  to  EN  13060  standard,  unwrapped  solid  loads,  solid 
porous loads, hollow A type loads, hollow B type loads, single and double wrapped loads can be 
sterilized in small steam sterilizers. 

NC  23B  Steam  Sterilizers,  come  with  5  operating  programs  for  sterilization  temperatures  of 
134°C and 121°C, 2 special programs, 1 drying program and 2 test programs. They can be used 
for  wrapped  or  unwrapped,  porous  textile  or  hollow  loads,  metal,  glass  and  rubber  materials 
that can be sterilized using pressurized and saturated water  vapor. NC 23B are not compatible 
with liquid sterilization. 

The steam is produced by the steam generator. The jacket system which  surrounds outside of 
the  chamber  provides  temperature  homogeneity.  Jacket  system  is  also  used  for  pre-heating 
which helps to decrease total sterilization cycle duration. 

All programs operate full-automatically without user’s interference. 

NC  23B  steam  sterilizers  have  several  safety  functions  with  its  integrated  safety  thermostat, 
safety valve and surface thermostat in addition to the safety features of the control system (i.e. 
high pressure, high temperature). 

NC  23B  Steam  Sterilizers  are  designed  and  manufactured  in  accordance  with  international 
directives and EN 13060, EN 61010-1, EN 61010-2-040, EN 60601-1-6, EN 61326-1, EN 62304, 
EN  62366-1,  EN  ISO  14971,  EN  50419,  EN  13445  and  EN  ISO  15223-1  standards  under  the 
supervision of total quality management systems ISO 9001 and ISO 13485. 

This device is in compliance with WEEE Regulation. 

Do not operate the instrument for purposes other than main purpose. 

This instrument is only to be used by authorized people after the user manual has been read. 

If the warnings mentioned in this manual are not considered, NUVE will not be responsible 
from their results.

 

 

Summary of Contents for NC 23B

Page 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A NC 23B BENCH TOP STEAM STERILIZERS USER S MANUAL 1984 Z14 K26 246 Rev No 8 Rev Date 11 2017...

Page 2: ...te damaged instrument 4 do not operate instrument with a damaged cable 5 do not move instrument during operation In case of a problem contact your N ve agent for an authorized service or maintenance T...

Page 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Page 4: ...MAINS SUPPLY 10 5 5 POSITIONING 10 5 6 GENERAL PRESENTATION 11 5 7 CONTROL PANEL 12 5 8 PRIOR TO OPERATION 14 5 8 1 CONNECTION TO MAINS 14 5 8 2 CONNECTION TO WATER SUPPLY 14 5 8 3 ATTACHING MICROBIOL...

Page 5: ...DATE 27 7 3 7 LANGUAGE 27 7 3 8 PASSWORD 27 7 3 9 SMS 28 7 3 10 E MAIL 29 7 4 C BAR 30 7 5 SERVICE 30 8 PERIODIC MAINTENANCE AND CLEANING 30 8 1 PERIODIC MAINTENANCE 30 8 2 PERIODIC CONTROL 30 8 3 CL...

Page 6: ...materials that can be sterilized using pressurized and saturated water vapor NC 23B are not compatible with liquid sterilization The steam is produced by the steam generator The jacket system which s...

Page 7: ...rs Pt 100 Chamber Material 304 Stainless Steel Steam Generator Material 304 Stainless Steel Chamber Volume Liters 23 Power Supply 230 V 50 60 Hz Power Consumption 2500W Control system N SmArtTM Progra...

Page 8: ...have a boiling point of 100 C at sea level at 760 mmHg 1 atm atmospheric pressure Prior to sterilization items which would be sterilized should be cleaned and disinfected The load which would be steri...

Page 9: ...A TOSB stanbul Tuzla Organize Sanayi B lgesi Tepe ren Mevkii 34957 Tuzla stanbul TURKEY Symbol in the operating instructions This symbol refers to important circumstances Labels on the product Earthed...

Page 10: ...z Please make sure that the supplied mains matches the required power ratings which are written on the name of plate of the instrument located at the back of the instrument Always plug in the instrume...

Page 11: ...ent 3 Fuses Main fuses on the phase neutral lines 4 On Off Switch Puts the unit on and off power 5 Thermal Printer After operation is completed printer provides hard copy of sterilization parameters 6...

Page 12: ...12 Water Tank Lid Has to be opened to fill up the water tank 5 7 CONTROL PANEL Figure 4 The functions of F1 F2 F3 and F4 keys depend on the meaning of the corresponding symbol appearing on the displa...

Page 13: ...ol denotes the left key It appears on the Main menu Programs and Graph screen On Main menu and Programs menu previous menu item is chosen when the key corresponding to this symbol is pushed On graph s...

Page 14: ...MICROBIOLOGICAL FILTER Figure 5 Mount the microbiological filter Figure 5 provided with the device to the top of the device where the filter fitting point is 5 8 4 STERILIZATION PROGRAMS 23B steam ste...

Page 15: ...d materials as metals on which water do not accumulate LOADING WAY MATERIAL TYPE PROGRAM NAME Program Temperature Wrapped Load Unwrapped Load Textile Solid Material 134 C Max 1 8kg Max 5kg 134 C Max 1...

Page 16: ...any case to be considered indicative in that the date of preservation is influenced by various factors as the environmental microbic level the granulometry of environmental dusts that act as carriers...

Page 17: ...g rules should be observed Instruments and sterilization containers should be placed at the bottom Textiles should always be placed at the top If sterilization bags and instruments are loaded together...

Page 18: ...other that they remain separate Pay attention to the guidance of the manufacturer of the instrument Where appropriate instruments should be disassembled before placing them in the autoclave as this wi...

Page 19: ...each sufficient steam temperature before sterilization phase STERILIZATION The chamber temperature is kept at the required sterilization degree all through the sterilization phase STEAM DISCHARGE The...

Page 20: ...color changes to blue and enter F4 key is used to access the selected menu item Password query screen on the left appears while accessing to Programs Special Programs Test Programs and Settings Enter...

Page 21: ...he top left corner and related sterilization cycle phase time appears at the top right line Push C Bar key F2 on the working screen to see the temperature and pressure sensor values which are placed i...

Page 22: ...ams page on the left Use increase F1 and decrease F2 keys to select the test program to operate Working screen appears when enter F4 key is pushed on the selected test program Push start F4 key on the...

Page 23: ...rous factors as ambient microbic level size of the dust particles ambient conditions of temperature pressure and humidity as well as the degree of handling of the sterilized materials themselves The m...

Page 24: ...name date time and sterilization result Recent Cycles displays more detailed information regarding cycle than Older Cycles Error Record submenu is selected on the memory page by the value increase F1...

Page 25: ...ings page on the left Use increase F1 and decrease F2 keys to select a submenu Working menu is selected from the main menu by using left F1 and right F2 keys Push enter F4 key to access settings page...

Page 26: ...ted as alphanumeric via Thermal Printer that is optional accessories Printer should be selected as ON from printer settings page to take printout via Thermal Printer 7 3 4 BUZZER Alarm Repeat Time sub...

Page 27: ...e left appears when enter key F4 is pushed The language of the control panel can be set as Turkish English French Russian or Spanish Push the value increase F1 and decrease F2 keys to select the langu...

Page 28: ...umber Each digit of phone numbers is entered one by one The first digit of the phone number is entered by pushing value increase F1 and decrease keys F2 and tab key F3 is pushed to enter the second di...

Page 29: ...ne by one The first character is entered by pushing value increase F1 and decrease keys F2 and tab key F3 is pushed to enter the next character The second character is entered by pushing value increas...

Page 30: ...ight F2 keys nad push enter F4 key to access Service menu is password protected and it is for authorized technical staff 8 PERIODIC MAINTENANCE AND CLEANING 8 1 PERIODIC MAINTENANCE Cleaning and perio...

Page 31: ...ectiveness loading unloading 6 months Annual 13 Observe the closing device for excessive wear 5 years 14 All safety valves exposed to direct steam pressure must be checked 5 years Safety tests pressur...

Page 32: ...is on The plug is plugged in properly The plug is not defective The mains supply is present 10 1 ERROR CODES Error Codes may appear immediately after the sterilizer is turned on or following a time la...

Page 33: ...urized air is released Thus the filter on the air release line should be sterilized by running quick program after power failure In case of power failure during sterilization phase chamber temperature...

Page 34: ...is started when the chamber is hot Vacuum Test Program should be started before sterilization phase and the chamber is cold If an error occurs please contact with an authorized N ve agent to seek tech...

Page 35: ...Figure 2 Ensure that power led is turned on See Figure 15 Power led is on when energy is supplied to the GSM module SMS led starts to flash while the module sending SMS Connect the antenna cable to an...

Page 36: ...36 12 ELECTRICAL CIRCUIT DIAGRAM...