24
When you select
“All Cycles”
on memory
page by using increase (F1) and decrease
(F2) keys and push enter (F4) key, the page
on the left appears.
When “Recent Cycles” is selected on all
cycles page, date query screen for
requested program appears on the screen.
After entering the date and time by using
increase (F1) and decrease (F2) keys and
pushing enter (F4) key, a page consisting of
records of cycles appears as on the left.
Choose a cycle by using increase (F1) and
decrease (F2) keys and push enter (F4) key
to see detailed record of cycle. Push print
(F4) key to print out the record.
“Older Cycles” page contains program name, date, time and sterilization result. “Recent Cycles”
displays more detailed information regarding cycle than “Older Cycles”.
“Error Record” submenu is selected on the
memory page by the value increase (F1) and
decrease (F2) keys. The page on the left
appears when enter key (F4) is pushed.
Failures are listed from the current date to
the earlier. Push the value increase (F1) and
decrease (F2) keys to pass the other page.
Push the backspace key (F4) in order to exit
the page.
Select “Cycle Info” submenu on memory
page by using value increase (F1) and
decrease (F2) keys and push enter (F4) key.
The page consisting of number of daily
cycles and number of total cycles appears.
Total number of cycles left to next
replacement of gasket and filter are also
shown on this page. Push the backspace key
(F4) in order to exit the page.