“Programs” menu is selected from the
main menu by using left (F1) and right (F2)
keys. Push enter (F4) key to access
programs page on the left. Use left (F1) and
right (F2) keys to select the program to
operate. Working screen appears when
enter (F4) key is pushed on the selected
program.
The screen on the left is working screen and
on working screen sterilization chamber
temperature and pressure values can be
monitored. Current date and time are
shown at the top line. Push enter (F4) key
to start the selected program. When
program starts to run, total sterilization
cycle time appears at the top left corner
and related sterilization cycle phase time
appears at the top right line.
Push “
°C / Bar
” key (F2) on the working
screen to see the temperature and pressure
sensor values which are placed in the
different parts of the sterilizer. Back key (F3)
is used to exit the screen. You may also
access this page by using “
°C / Bar
” sub
menu on main menu.
When the graph button (F3) is pushed during the operation, the graph page showing pressure
and temperature values appears on the screen.
6.2.1.
SPECIAL PROGRAMS
“Special Programs” menu is selected from
the main menu by using left (F1) and right
(F2) keys. Push enter (F4) key to access
special programs page on the left. Use tab
(F1) key to select the program to operate.
Working screen appears when enter (F4)