Cleaning shall be performed while chamber is cold.
9.
DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the
responsibility of the user to arrange proper disposal of the individual components.
All parts which may comprise potentially infectious materials have to be disinfected by suitable
validated procedures (autoclaving, chemical treatment) prior to disposal. Applicable local
regulations for disposal have to be carefully observed.
The instruments and electronic accessories (without batteries, power packs etc.) must be
disposed off according to the regulations for the disposal of electronic components.
Batteries, power packs and similar power source have to be dismounted from electric/electronic
parts and disposed off in accordance with applicable local regulations.
10.
TROUBLESHOOTING
If the device fails to operate, please check the followings:
The power switch is on;
The plug is plugged-in properly;
The plug is not defective;
The mains supply is present.
10.1.
ERROR CODES
Error Codes may appear immediately after the sterilizer is turned on or following a time lag
after the unit is turned on, before any program is started.
Error Codes may appear immediately after a program is started or during any program
execution.
These messages are accompanied by an alarm tone which can be shut down by the “stop”
button.
In case of any failure during a program run; the program is interrupted and the sterilizer either
releases steam or vents the chamber according to the pressure condition in the chamber.
In case of any failure before sterilization phase during a program run, the load would
not be sterile
since the sterilization phase is not started.
In case of any failure after sterilization phase during a program run, the load would be
sterile
. However, loads could be
wet
, since drying phase would be interrupted.
Failures which may be encountered during operation is listed below:
Error 01:
Vacuum Time Exceeded –
The chamber pressure does not drop to the required
vacuum value within preset duration.