11.0 Electrical/Environmental Requirements
11.1 Electrical (Supply Voltage Fluctuations Not to /- 10%)
*NU-543-300S
230 VAC,
50/60 Hz,
1 Phase,
8 Amps
*NU-543-400S
230 VAC,
50/60 Hz,
1 Phase,
10 Amps
*NU-543-500S
230 VAC,
50/60 Hz,
1 Phase,
11 Amps
*NU-543-600S
230 VAC,
50/60 Hz,
1 Phase,
11 Amps
*CE Certified
11.2 Operational Performance (for indoor use only)
Environment Temperature Range:
60
F-85
F (15
C - 30
C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to 31
C decreasing
linearly to 50% relative humidity at 40
C.
Environment Altitude:
6562 Feet (2000M) maximum
11.3 Light Exposure
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
11.4 Installation Category: 2.0
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example,
in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains
such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.
11.5 Pollution Degree: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
11.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting. USE
OF CHLORINATED OR HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL.
Equipment decontamination can be accomplished by non-condensing gas or vapor Paraformaldehyde, Hydrogen Peroxide
or Chlorine Oxide following NSF/ANSI 49, Annex G.
11.7 EMC Performance (classified for light industrial)
Emissions:
EN61326
Immunity:
EN61326
Class A equipment is intended for use in an industrial environment.
In the documentation for the user, a statement shall be included drawing
attention to the fact that there may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted
as well as radiated disturbances.
WARNING
!
Summary of Contents for LabGard ES NU-543-300S
Page 7: ...OM0239 Page 7 of 72 April 2021 ...
Page 9: ...OM0239 Page 9 of 72 April 2021 ...
Page 10: ...OM0239 Page 10 of 72 April 2021 ...
Page 11: ...OM0239 Page 11 of 72 April 2021 ...
Page 12: ...OM0239 Page 12 of 72 April 2021 ...
Page 16: ...OM0239 Page 16 of 72 April 2021 ...
Page 21: ...OM0239 Page 21 of 72 April 2021 ...
Page 35: ...OM0239 Page 35 of 72 April 2021 ...
Page 67: ...OM0239 Page 67 of 72 April 2021 ...
Page 68: ...OM0239 Page 68 of 72 April 2021 ...
Page 69: ...OM0239 Page 69 of 72 April 2021 ...
Page 70: ...OM0239 Page 70 of 72 April 2021 ...