OM0239
Page 42 of 72
April 2021
7.4 Sliding Window Replacement and Manual Adjustment
The sliding window replacement is accomplished by removing the front decorative panel, control center, and window glide
assemblies. The sliding window adjustment may be required due to everyday use over the life of the cabinet. Both window
glides are adjustable by a set screw and tension screw method. When adjusting the sliding window, be sure to verify
proper micro switch operation. If the sliding window is too loose, it will not properly activate the micro switches, thus
causing potential operational malfunctions to occur. If the window counter balance experiences a fault, window function
must be verified (the sliding window retention or ability to slow the rate of fall). See Sliding Window Retention Verification
in the Inspection Report).
7.5 Airflow Control System Setup and Calibration
7.5.1
General
The operation of the NU-543S cabinet requires that the setup and calibration procedures be performed in order to
certify or commission the cabinet for usage. The setup and calibration procedures performed
ONLY BY THE
CABINET CERTIFIER
ensure that cabinet's setpoints are verified and that the airflow monitor sensors are calibrated
to read the correct values. Press MENU to access Calibration/Service parameter.
To enter the service parameter menu, perform the following:
•
Press and hold
key, then press ENT
↑ and ↓
keys sequentially releasing the
key after the
3 key sequence.
The following service parameter display should appear
Navigation through the service parameter menu is the same as the operator menu.
•
Press ↑ on ↓ keys to select a menu item which will be highlighted inversely. Menu selection moved in a
round robin pattern.
•
Press ENT key on a selected menu item will either:
3)
Drill further down the menu tree
4)
Allow editing of the item value which will now be highlighted.
o
Press ↑ or ↓ key to change the selected value
o
Press ENT key to accept the change
o
Press
key to cancel the change
•
Press
key to back up one level or repeated times to return to normal run or standby mode.
7.5.2
Configuration Parameters
Configuration parameters identify both user desired and required functional control system characteristics. User
desired parameters includes language, flow units, temp units and logo. Requires functional parameters include
motor type and cabinet size which assures proper performance characteristics.
Parameter value choices:
Language:
English
, Spanish, German, French, Chinese, Japanese
Flow Units:
m/s
,
fpm
Temp Units:
⁰C
, ⁰F
Logo:
NuAire
Motor Type:
ECM AUTO
, EC FIXED, ECM FIXED
Cabinet Size: 3 ft.,
4 ft.
, 5 ft., 6 ft. Lam 3-4 ft., Lam 5-6 ft.
Bold
items represent default parameters
Service Parameters
Configuration
Setpoints
Calibration
Options
Configuration Parameters
Logo
NuAire
Motor Type:
ECM Auto
Cabinet Size
: 4 ft.
Configuration Parameters
Language
English
Flow Units:
m/s
Temp Units:
⁰C
More
Summary of Contents for LabGard ES NU-543-300S
Page 7: ...OM0239 Page 7 of 72 April 2021 ...
Page 9: ...OM0239 Page 9 of 72 April 2021 ...
Page 10: ...OM0239 Page 10 of 72 April 2021 ...
Page 11: ...OM0239 Page 11 of 72 April 2021 ...
Page 12: ...OM0239 Page 12 of 72 April 2021 ...
Page 16: ...OM0239 Page 16 of 72 April 2021 ...
Page 21: ...OM0239 Page 21 of 72 April 2021 ...
Page 35: ...OM0239 Page 35 of 72 April 2021 ...
Page 67: ...OM0239 Page 67 of 72 April 2021 ...
Page 68: ...OM0239 Page 68 of 72 April 2021 ...
Page 69: ...OM0239 Page 69 of 72 April 2021 ...
Page 70: ...OM0239 Page 70 of 72 April 2021 ...