OM0239
Page 58 of 72
April 2021
8.2.1 Sync Outputs with Active Blower
Sync outputs with active blower allows for the listed functions to be turned on and off automatically when the
blower is actively running. If the blower is pending (blinking LED above blower switch), these listed functions will not
turn on if selected as active. ONLY when the blower is actively running will these functions turn on.
Fan Relay - Normally the fan relay will activate when the blower switch is pressed. Blower can either be actively
running or pending. If the fan relay sync is active the blower must be actively running for the relay to
change state.
Fluorescent Light
–
Normally the fluorescent light is turned on via the fluorescent light key. If the fluorescent light
sync is active; the blower must be actively running for the fluorescent light to turn on.
15Volt DC Output
–
Normally the 15 Volt DC output located on the control board is on when power is applied to the
FlowGard™
system. If the 15 volt DC output is active, the blower must be actively running for
the 15 volt DC output to turn on.
Accessory Outlet
–
Normally the accessory outlet is on all the time. If the accessory outlet sync is active, the blower
must be actively running for the accessary outlet to turn on.
Outlet Power
–
Normally the outlet power is turned on via the outlet key. If the outlet power sync is active, the
blower must be actively running for the outlet power to turn on.
8.2.2 Function Options
Disable Intelliflow
–
If this function is active. All airflow display and alarm functions are turned off.
Allow UV light anytime
–
Normally the UV light is interlocked with the window being in the closed position. For
service purposes only, if this function is active, the UV light may be turned on at any
window height.
Note:
IN addition to the Allow UV light anytime system function, there is a double redundant UV light window
interlock relay. To override the UV light window interlock relay, the relay itself must also be shorted
(see electrical schematic for reference).
Alarm Active in Warm Up
–
Normally the audible and visual alarm is not active during the blower warm up period.
If this function is active, both an audible and visual alarm will occur during the blower
warm up period. The audible alarm may be silenced, but the visual alarm will remain
active until the warm up period is complete.
Airflow Alarm Silence
–
Normally airflow alarms that occur can be temporarily silenced (i.e. default audible silence
time is 15 minutes) by pressing the alarm silence key. If this function is turned off, the audible
alarm cannot be silenced.
Sync with Active Blower
Fan Relay: No
Fluorescent Light: No
15 Volt DC Output
No
Accessary Outlet:
No
Outlet Power: No
Function Options
Disable Intelliflow:
No
Allow UV Light Any time: No
Alarm active in WarmUp: No
Airflow Alarm Silence: Yes
Summary of Contents for LabGard ES NU-543-300S
Page 7: ...OM0239 Page 7 of 72 April 2021 ...
Page 9: ...OM0239 Page 9 of 72 April 2021 ...
Page 10: ...OM0239 Page 10 of 72 April 2021 ...
Page 11: ...OM0239 Page 11 of 72 April 2021 ...
Page 12: ...OM0239 Page 12 of 72 April 2021 ...
Page 16: ...OM0239 Page 16 of 72 April 2021 ...
Page 21: ...OM0239 Page 21 of 72 April 2021 ...
Page 35: ...OM0239 Page 35 of 72 April 2021 ...
Page 67: ...OM0239 Page 67 of 72 April 2021 ...
Page 68: ...OM0239 Page 68 of 72 April 2021 ...
Page 69: ...OM0239 Page 69 of 72 April 2021 ...
Page 70: ...OM0239 Page 70 of 72 April 2021 ...