SpO2 alarm preset
limits
Upper alarm limit
1%~100%
Lower alarm limit
0%~99%
SpO2 alarm preset
accuracy
±1%
SpO2 alerting signal
generates a delay
No delay
SpO2 value refresh
period
1s/time
SpO2 value refresh
delay
<
10s
Average period
Low sensitivity
7
~
8s
Intermediate sensitivity
4
~
6s
Advanced sensitivity
2
~
3s
Alarm condition delay
period
Low sensitivity
<
8s
Intermediate sensitivity
<
6s
Advanced sensitivity
<
3s
Alarm sign generates
delay period
0s
PR
Measuring range
30~254bpm
Resolution
1%
Accuracy
±2% or ±2bpm