![Northern Meditec Taurus E Operator'S Manual Download Page 65](http://html1.mh-extra.com/html/northern-meditec/taurus-e/taurus-e_operators-manual_1708589065.webp)
文件编号:
NR-TP-0715-008
版本:
1.0
- 53 -
1.
Connect the cuff to the NIBP pore of the monitor
2.
Wrap the cuff on the cylinder of appropriate size.
3.
Set the patient type to adults.
4.
Select
[Menu] Smart Hotkey → [Main Menu];
5.
Select [User Maintenance >>] → enter the password and confirm → [User Maintenance]
menu;
6.
Select [Module Maintenance >>] → [Module Maintenance] menu;
7.
Select [NIBP >>] → [NIBP Maintenance] menu, and select [Leak Test] for leakage test.
8.
After 20s, the system will automatically open the bleeder valve, and the leak test finishes.
9.
If the NIBP parameter area displays [Leakage test Stopped], the system does
not leak. If it displays [Cuff leak], it indicates that the air passage leaks. At this time, the
operator should check if the entire connection is loose, and test for leak once again when
the connection is correct. If there is error prompt still, please contact the manufacturer for
repair.
Fig. 11-2 NIBP Leakage Test Connection Diagram
10.9
Calibration
Users can not calibrate NIBP. If calibration is required, please contact your service
representative. Cuff pressure sensor should be checked and calibrated at least once every two
years by qualified professional service personnel.
10.10
Clean and disinfection
If necessary, NIBP cuff and NIBP extension tube can be cleaned and disinfected together
without separated
10.10.1
Cleaning
1.
Prepare enzyme cleaning agent, distilled water and 10% solvent, respectively in
different spray bottle.