![Northern Meditec Taurus E Operator'S Manual Download Page 20](http://html1.mh-extra.com/html/northern-meditec/taurus-e/taurus-e_operators-manual_1708589020.webp)
文件编号:
NR-TP-0715-008
版本:
1.0
- 8 -
monitor makes a “tick” sound, the warning system of the monitor can work normally.
Notes
If any fatal error occurs during self-test, the system will alarm.
Check all available monitoring functions to ensure that the monitor functions
properly.
If the monitor integrates a battery, charge the battery after each use to secure
sufficient power.
2.4.3
Connecting the Sensor
Connect the required sensor to the monitor and the monitoring position on the patient.
Note
For proper connections and related requirements for a variety of sensors, see
Chapters 7-12.
2.4.4
Starting Monitoring
Start monitoring in the following steps:
Check if the patient cable and the sensor are connected properly.
Check if the settings of the monitor are correct, such as: patient category [Patient Cat.]
For the details of parameter measurement or monitoring, see the appropriate section.
2.5
Turning off
Please turn off the monitor in the following steps:
Disconnect the cables and sensors connected to the patient.
Press and hold the power switch for two seconds to pop up the 5sec countdown window,
and the monitor turns off in five seconds.
Caution