If the instrument function has any sign of damage, do not use this monitor for any patient
monitoring. Please contact the hospital's professional maintenance personnel or our customer
service personnel.
All checks that require disassembling the instrument must be performed by qualified service
personnel. Safety and maintenance checks may also be carried out by the Company's
personnel.
15.2
Viewing Software Version Info
You can view the software version through the following steps:
Select
[Menu] Smart Hotkey→[Main Menu];
Select [Monitor Information >>] → [Monitor Info] menu;
[Monitor Info] menu displays the software version information of the monitor.
The software version of the monitor: V1.1.7
15.3
Maintenance Plan
The following tasks can only be done by qualified service personnel of Northern. When the
following maintenance is needed, please contact your service representative. Before testing
or maintenance, clean and disinfect the device.
Inspection / Maintenance Item
Frequency
Check the safety according to
IEC 60601-1
At least once every two years, after replacing the power
supply or the monitor falls down.
Check all monitoring or
measuring functions not listed
At least once every two years, or when you suspect that the
measured value is not accurate.
NIBP leakage test
At least once every two years, or follow hospital regulations
NIBP calibration
At least once every two years, or follow hospital regulations