Fig. 11-1 NIBP Parameter Area
10.6
Inflation Pressure
If necessary, you can manually set the initial cuff inflation pressure as follows:
Select the NIBP parameter area → [NIBP Setup] menu;
Select [Pre-Infl Press], and set the appropriate cuff pressure value.
10.7
Resetting
Select NIBP parameter area → [NIBP Setup] menu → Select [Reset], and restore the
inflation pressure of the blood pressure pump to currently configured initial settings. When
the blood pressure pump is not working properly, but no warning is given, you can reset the
blood pressure pump, and automatically restores the blood pressure pump.
10.8
Leakage Test
The purpose of leakage test is to detect if the sealing of the air passage is in good condition.
If the leakage test passes, the NIBP parameter area displays [Leakage test Stopped]. If not
passed, the NIBP parameter area displays [Cuff leak] message. NIBP leakage test shall be at
least once every two years or when you think that the reading is not accurate.
Prepare the following materials before the test:
Adult cuff: one
Inflation tube: one
Cylinder of appropriate size: one
Leakage test process