User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
23
-
Screen Description:
“SpO2
”:
SpO
2
alarm switch; “” indicates SpO2
alarm is on; “
”
indicates SpO2
alarm is off.
“SpO2 Hi”:
high limit of SpO2
alarm; range: “1~100”.
“SpO2 Lo”:
low limit of SpO2
alarm; range: “0~99”.
“PR
”:
pulse rate alarm switch; “” indicates PR alarm is on; “
” indicates
PR
alarm is off.
“PR Hi”:
high limit of PR
alarm; range: “22~250”.
“PR Lo”:
low limit of SpO2
alarm; range: “0~248”.
“PI display”:
“on” means PI display is enable; “off”
means PI display is
disable.
Operation Instructions
1.
Press “▲” key or “▼” key to move cursor to select parameter.
2. Press “■” key to confirm and active this parameter setting.
3. Press “▲” key or “▼” again to adjust or modify parameter value.
4. Press “■”key again to confirm and save the setting.
5. Press
“
” key to return to upper level screen.
4.4.2 NIBP Setup
Figure 4.7 NIBP Setup Screen
“unit”: unit of the blood pressure value;
“mmHg” or “kPa” can be selected. Conversion: 1kPa=7.5mmHg.
“SYS
”: systolic pressure alarm switch; “ ” indicates systolic pressure alarm is on;