User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
16
-
compartment.
5. Pull the paper out of paper slot on the shield.
6. Move the tab of the rubber roller lock 90° downwards to lock it.
7. Put the shield back in position and secure it.
Operation procedures for taking out printing paper roll:
1~2 steps are the same with the 1~2 steps mentioned above for loading printing paper.
3. Roll the loading roller anti-clockwise and pull the paper out.
4~5 steps are the same with the 6~7 steps mentioned above for loading printing paper.
Figure 3.6 Loading and taking out printing paper
3.3.4 Battery Installation
1. Ensure that the monitor is not connected to AC power supply and the monitor is turned
off.
2. Open the battery cover and place the battery in the direction as shown in Fig. 3.7 to insert
the battery into any one of battery compartments. Do not insert battery with their
polarities reversed.
3. Move the battery baffle to secure battery.
4. Close the battery cover.
Note: you can install optional one more battery to prolong the working time.