Figure 3. 5 Neonate SpO2 sensor placement
Safety Introductions for SpO2 Monitoring
Continuous use of finger clip SpO2 sensor may result in discomfort or pain, especially
for those patients with microcirculatory problem. It is recommended that the sensor
should NOT be applied to the same finger for over two hours.
SpO2 measuring position must be examined more carefully for some special patient.
Do NOT place the SpO2 sensor on the finger with edema or fragile tissue.
Please do not allow the cable to be twisted or bended.
Do NOT put the SpO2 sensor and pressure cuff on the same limb, otherwise the NIBP
measuring will affect SpO2 measuring and cause the alarm error.
Using nail polisher or other cosmetic product on the nail may affect the accuracy of
measurement.
The fingernail should be of normal length.
Do NOT use the damaged SpO2 sensor.
The SpO2 sensor can not be immerged into water, liquor or cleanser completely,
because the sensor has no capability of waterproofness.
3.3.3 Loading printing paper
Operation procedures for loading printing paper:
1. Press both “OPEN” notches with force on printer shield with two thumbs to open it.
2. Move the tab of rubber roller lock at the left 90°upwards to unlock it, refer to the
following figure with mark
①
.
3. Cut one end of the paper into triangle, and load the paper from the underside of the rubber
roller.
4. Turn the roller clockwise to get the paper rolled, and put the paper roll into the