the function. Use ESC to exit.
All errors are written into the error memory
(see “27 Operation Mode (RUN)” / “Error
message”).
>ON delay<
The “ON” event in case of reaching the limit
value can be delayed by up to 9999 seconds
max. The relay will not energise before the
time set is expired and the limit value is pre-
sent yet.
If the value falls below the limit threshold for
a moment the cycle will begin anew.
>OFF delay<
The “OFF” event in case of reaching the limit
value can be delayed by up to 9999 seconds
max. The relay will not de-energise before
the time set is expired and the limit value is
present yet.
If the value falls below the limit threshold for
a moment the cycle will begin anew.
>Name<
The relay output name may consist of four
characters max. which will be indicated in
main menu and in overview menu. Set the
name as described in chapter “29.1 Parame-
ter Menu “Measurement Place””.