Parameter Setting
NFP - rev. 04 / 23.05.2018
page 63
30.6 I/O sub menu “v-Sensor”
This screen indicates sensor number and software version, measured velocity, investigated
Reynolds correction, transmitting frequencies and velocity of sound.
This is primarily for service purposes.
1
Software version of sensor
2
Creation date of sensor software
3
Calculated average velocity
4
Exponent of Reynolds function
5
Impulse repeat frequency
6
Measured medium temperature
7
Velocity of sound resulting from medium temperature
Fig. 30-7
v-Sensor
Use >left< and >right< arrow keys to browse through more screens.
The velocity graph represents the velocity distribution of individual gates within the pipeline.
This graph enables to assess the prevailing hydraulic conditions.
Fig. 30-8
Velocity graph
The velocity distribution diagram combines all measured flow velocities in all gates and rep-
resents their distribution within frequency groups.
The shape of this distribution may provide information on asymmetric hydraulic conditions,
vorticity or similar.
Fig. 30-9
Velocity distribution diagram
The last screen provides information on amplification, cable noise and various evaluation re-
sults between transmitter and sensor.
Summary of Contents for NFP
Page 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...
Page 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...
Page 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...
Page 85: ...Approvals and certificates NFP rev 04 23 05 2018 page 85...
Page 86: ...Instruction Manual NIVUS Full Pipe page 86 NFP rev 04 23 05 2018...
Page 87: ...Approvals and certificates NFP rev 04 23 05 2018 page 87...