Parameter Setting
NFP - rev. 04 / 23.05.2018
page 45
>Time format<
12- or 24-hours format
Fig. 28-5
System time submenus
>Modif. totalizer<
This menu allows to newly set the totalizer indicated on the main
screen. This function is used if the transmitter has been replaced and
the new transmitter is to show the same total value.
-
Specify new total value
-
Confirm with Enter key
-
Enter system PIN and confirm
-
New total value is shown in the main screen
Fig. 28-6
Modification of totalizer
>Modify PIN<
>System-PIN<
The system PIN is the transmitter password
and for parameter modifications.
Default system PIN: “2718”.
NIVUS recommend changing the PIN to pro-
tect the system from unauthorised access.
The PIN is arbitrarily selectable (max. six
digits).
Tip:
For your own safety we recommend to share