>ON point<
Defines the “ON” point for the selected limit
value.
>OFF point<
Defines the “OFF” point for the selected limit
value.
>ON delay<
The “ON” event in case of reaching the limit
value can be delayed by up to 9999 seconds
max. The relay will not energise before the
time set is expired and the limit value is pre-
sent yet.
If the value falls below the limit threshold for
a moment the cycle will begin anew.
>OFF delay<
The “OFF” event in case of reaching the limit
value can be delayed by up to 9999 seconds
max. The relay will not de-energise before
the time set is expired and the limit value is
present yet.
If the value falls below the limit threshold for
a moment the cycle will begin anew.
>Name<
The relay output name may consist of four
characters max. which will be indicated in
main menu and in overview menu. Set the
name as described in chapter “29.1 Parame-
ter Menu “Measurement Place””.
Fig. 29-14 Limits sub menu
>Vel. limit relay<
Relay will energise if a velocity limit value (to be set) has been ex-
ceeded and will de-energise if velocity falls below a second limit value
to be set).
>Relay mode<, >ON
point<, >OFF point<,
>ON delay<, >OFF
delay<, >Name<
See >Flow limit relay<
>Temp. limit re-
lay<
Relay will energise if a medium temperature limit value (to be set) has
been exceeded and will de-energise if the Medium temperature falls
below a second limit value.
>Relay mode<, >ON
point<, >OFF point<,
See >Flow limit relay<
Summary of Contents for NFP
Page 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...
Page 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...
Page 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...
Page 85: ...Approvals and certificates NFP rev 04 23 05 2018 page 85...
Page 86: ...Instruction Manual NIVUS Full Pipe page 86 NFP rev 04 23 05 2018...
Page 87: ...Approvals and certificates NFP rev 04 23 05 2018 page 87...