prw2101a0600p_01
12 / 105
2.
TECHNICAL DATA
T
YPE
P
-
2
-
Transmitter Power Supply
30 V DC / 60 mA (For Ex Version: 25 V DC / 22 mA)
Display
128 x 64 dot-matrix
Analogue
Max. 2 galvanically isolated 4 … 20 mA outputs,
max load of 500 ohm, with over-voltage protection
Relay
Max. 5 SPDT
,
250 V AC, AC1. 5 A
RS 485 interface
Galvanically isolated, HART / MODBUS protocol
HART
Output signal level: 0.5 ±0.1 V
pp
trapezoid 1200 / 2200 Hz
Minimum level of Input signal: 50 mV
pp
Inbuilt sensor resistance: 250 ohm.
Outputs
Logger
Capacity: FLASH = 65000 entry; SD card (max 2 GB) = depends on the card!
T
YPE
P
-
2
-
Power Supply, Relays, analogue 4 … 20 mA
0.5 … 2.5 mm
2
wire cross-section
RS 485 interface
Shielded, twisted cable pair with a cross section of 0.5 … 2.5 mm
2
Connecting
cables
HART line
Below1500 m shielded, twisted cable pair with a min.cross section of
0.5mm
Above 1500 m shielded, twisted cable pair, with a min. cross section of
0.8mm
max. resistance : 75 ohm, max.capacitance : 200 nF
Number of powered transmitters
Max. 15 non-Ex (max. 4 Ex) transmitters
Power supply /
Power consumption /
max. power supply
85 … 255 V AC 50 … 60 Hz / 12 VA / 255 V
eff
11.4 … 28 V AC 50 … 60 Hz / 12 VA / 28 V
eff
11.4 … 40 V DC / 11 W / 40 V DC
Fuse
85 … 255 V AC 50 … 60 Hz T400 mA
11.4 … 28 V AC 50 … 60 Hz és 11.4 … 40 V DC T1A
Housing material
Polycarbonate (PC)
Mounting
Wall mounted
Ambient Temperature
-20 °C … +50 °C
Ingress protection
IP 65
Ex marking
II (1) G [Ex ia Ga] IIB
Intrinsically safe output limitation data
U
o
= 30 V I
o
= 140 mA P
o
= 1 W
L
o
= 4 mH C
o
= 200 nF
Electrical protection
Class I / III
Mass
0.9 kg
Summary of Contents for MultiCONT