BKI 11 ATEX 0012 X
le00100a0600p_05
51 / 60
Upper value of the simulation
Parameter: P87
Menu path:
SERVICE / SIMULATION / UPPER VALUE
Default value:
pH measurement
ORP measurement
DO measurement
Description:
Highest value of the simulation
14 pH
1000 mV
8.24 ppm
6.4.5.8
Temperature simulation
This function helps the user to check the outputs and the additional processing instruments connected to the output.
To start simulation the instrument must return to Measuring mode.
Temperature simulation mode
Parameter: P84:
b
Menu path:
SERVICE / TEMP. SIMULATION / MODE
Default value:
OFF
Description:
Temperature simulation mode:
OFF
No temperature simulation
FIX VALUE
Simulation of a fixed lower range temperature value
TRIANGLE WAVE
The simulated value changes linearly between the set low
and high values with an adjustable cycle time