32 / 60
le00100a0600p_05
BKI 11 ATEX 0012 X
PLACEMENT
The instrument should be mounted within a max. 45° angle, compared to vertical position.
Types mounted with 4xphekl112seph / 4xorrheklseor electrode may be positioned
horizontally as well!
If the unit is installed using a fitting piece the user has to assure that before and after the fitting piece
there is a straight pipe with the appropriate length to minimalize sedimentation and turbulency.
The influent section has to have min. 10xDN length while the effluent section has to have min. 3xDN
length.
It is advised to create a by-pass line so during the service of the instrument the electrode is kept wet
or a calibration can be done without stopping the process.
>10 DN
>3 DN
LIQUID CONTINUITY
Near the electrode (sensor), the pipeline should always be filled with liquid: