36 / 60
le00100a0600p_05
BKI 11 ATEX 0012 X
6.
PROGRAMMING
The AnaCONT instruments can be programmed with the SAP-300 display unit.
Without SAP-300 the following LEDs can be seen:
RELAY
Indicates the energised state of the relay (C2 closed, C1 open)
VALID
Indicates the measuring capability of the unit. If lit, the input signal is stabilised.
If blinking,the input signal is changing.
COM
Indicates digital (HART) communication. During Remote programming LEDs are
continuously illuminated.
When SAP-300 display is present, LEDs are not visible; functions are taken over by the display. In default state SAP
display shows the primary measured value (which the output current is calculated from). Programming is supported by
a text-based menu. Navigation is done by using the
E
/ / / buttons.
The instrument can operate without the SAP-300 module as well. Programming can be done only by using SAP-300
display module locally or by HART (REMOTE PROGRAMMING) communication system remotely.
6.1.
T
HE
SAP-300
DISPLAY UNIT
The SAP-300 is a 64x128 dot-matrix LCD display which can be plugged into the transmitter. (Universal – usable in
other NIVELCO devices as well – provided that the system software supports SAP-300.)
Warning!
The SAP-300 module is based on LCD technology, so please make sure it is not exposed to permanent heat or direct
sunlight, in order to avoid damage of the display unit. If the instrument cannot be protected against direct sunlight or
high temperature that is beyond the standard operating temperature range of the SAP-300, please do not leave the
SAP display in the instrument.