6.4.2.
Analogue output
6.4.2.1
Output current mode
Parameter: P12:
b
Menu path:
OUTPUT SETUP / ANALOG OUTPUT / CURRENT MODE
Default value:
AUTO
Description:
Transmission mode of the current output [AUTO, MANUAL]
AUTO
The output current is calculated from the measured value, output is
active.
MANUAL
The output current is fixed at a constant (set) value (see
6.4.2.5
). In
this mode the setting of the error current (see
6.4.2.4
) is irrelevant.
The set (current) value overwrites the 4mA output of HART
multidrop mode!
6.4.2.2
Output current value 4mA
Parameter: P10
Menu path:
OUTPUT SETUP / ANALOG OUTPUT / 4mA VALUE
Default value:
pH measurement
ORP measurement
DO measurement
Description:
Measured value assigned to 4mA
0 pH
- 1000 mV
0 ppm
6.4.2.3
Output current value 20 mA
Parameter: P11
Menu path:
OUTPUT SETUP / ANALOG OUTPUT / 20mA VALUE
Default value:
pH measurement
ORP measurement
DO measurement
Description:
Measured value assigned to 20mA
14 pH
1000 mV
8.24 ppm
6.4.2.4
Output current error mode
Parameter: P12:
a
Menu path:
OUTPUT SETUP / ANALOG OUTPUT / ERROR MODE
Default value:
HOLD
Description:
Error indication by the current output
HOLD
Error indication has no effect on the output current.
3.8 mA
Error indication: the output current gets 3.8mA.
22 mA
Error indication: the output current gets 22mA
Attention!
Also see chapter
6.4.2.1
!