Information displays:
Press button to cycle between the information displays.
1.
The general information display (DEV. INFO): overall
running time (OV. RUN TIME), run time after power on
(RUN TIME), type of interface (INTERFACE), relay
(RELAY) and logger (LOGGER) indication.
2.
Sensor information: time passed after last sensor
change, slope and offset after sensor change, time
passed after last calibration, slope and offset at last
calibration, minimum / maximum temperature of the
sensor.
In case of DO measurement in addition to these:
atmospheric pressure (PRESSURE) and salinity
(SALINITY).
3.
Relay information: number of switchings, total energised
time (C2 closed) (only if there is a relay in the unit).
The informative display switches back to main screen after 30
seconds.
By pressing the button the user can get back to main screen at
any time.