42 / 60
le00100a0600p_05
BKI 11 ATEX 0012 X
6.4.
P
ROGRAMMABLE FEATURES DESCRIPTION
6.4.1.
Basic measurement settings
6.4.1.1
Measuring unit (Only for DO measurement)
Parameter: P02:
b
Menu path:
BASIC SETUP / MEASURING UNIT
Default value:
Dimension of PV: primary value.
ppm
ppm
mg/l
%sat
Description:
Primary values are the same in case of displaying ppm and mg/l, only
displayed and transmitted dimension changes. Large number on the screen
follows the dimension of ppm and mg/l in the setting of ppm and mg/l.
In case of choosing %sat, large numbered primary value indicates the oxygen
saturation.
6.4.1.2
Damping time
Parameter: P20
Menu path:
BASIC SETUP / DAMPING TIME
Default value:
pH
measurement
ORP
measurement
DO
measurement
Description:
Damping time is used to damp the unwanted fluctuations of the output and
display.
If the measured value changes rapidly the new value will settle with 1%
accuracy after this set time
2 sec
2 sec
2 sec
6.4.1.3
Temperature compensation mode
Parameter: P06:
a
Menu path:
BASIC SETUP / TEMP. COMPENSATION / MODE
Default value:
AUTO
Description:
Mode of the temperature compensation
AUTO
Compensation is done with the value measured by the temp. sensor.
MANUAL Compensation is done with a predefined temperature value
(see
6.4.1.5
)
OFF No
compensation