50 / 60
le00100a0600p_05
BKI 11 ATEX 0012 X
6.4.5.7
Simulation
This function helps the user to check the outputs and the additional processing instruments connected to the output. (Irrespectively of the instruments)
To start simulation the instrument must return to Measuring mode.
Simulation mode
Parameter: P84:
a
Menu path:
SERVICE / SIMULATION / MODE
Default value:
OFF
Description: Simulation
mode:
OFF
No
simulation
FIX VALUE
Its value is set according to the lower range value of pH / ORP
simulation
TRIANGLE WAVE
The simulated value changes linearly between the set low and
high values with an adjustable cycle time.
PV: Primary Value
SQUARE WAVE
The simulated value jumps between the set low and high values
with an adjustable cycle time.
Simulation cycle
Parameter: P85
Menu path:
SERVICE / SIMULATION / TIME
Default value:
Description:
Cycle time of the simulation
60 sec
Bottom value of the simulation
Parameter: P86
Menu path:
SERVICE / SIMULATION / BOTTOM VALUE
Default value:
pH measurement
ORP measurement
DO measurement
Description:
Lowest value of the simulation
0 pH
- 1000 mV
0 ppm