E-17
6
Install the remaining
three filter cubes to the
fluorescence attachment
in the same way and
return the filter cube
fixing plate to its
original place. Insert
the excitation plates in
the order of installation
starting from the rear to
the front.
NOTE: Be sure to install four filter cubes. If filter cubes needed for
observation is less than four, use the dia-filter cube supplied
with the attachment so that the total of four filter cubes are
installed.
7
Return the front cover to its original place.
7
When performing photomicrography or TV observation:
1
Remove the cap on the fluorescence attachment port mount by
loosening the clamp screw with the hexagonal wrench.
2
Mount the fluorescence attachment port (P-FLAP or P-FLAP2) and
tighten the clamp screw.
3
Remove the cap on the
fluorescence attachment
port and mount the
photomicrographic
equipment or high-
sensitive TV camera (via
the appropriate adapter).
Please refer to the
instruction manuals
supplied with the
photomicrographic
equipment or the TV
camera for their assembly and usage. Ask your nearest Nikon
representative for the best combination of adapter to be used in
between.
1
1
2
3
4
2
3
4
Installation order of filter cubes, and the
installation order of their excitation plates. (In
this figure, dia-filter cubes are installed at the
position 1 and 4.)
Fluorescence attachment port
(P-FLAP or P-FLAP2)
(This picture shows the P-FLAP.)
Assembly
II