E-16
6
Attach the filter cubes.
If your filter cube does not have any filters or dichroic mirror attached
(P-FLFC filter cube), first refer to the page E-22 and attach the filers to the
filter cube.
1
Push in the filter cube
switchover lever.
2
Remove the front cover
of the fluorescence
attachment by
loosening the clamp
screw with the
hexagonal screwdriver.
3
Remove the filer cube
fixing plate by
loosening the clamp
screw with the
hexagonal screwdriver.
4
Screw on the supplied
attachment tool to the
filter cube. Hold the
tool and insert the cube
into the groove on the
fluorescence
attachment. Screw off
the attachment tool.
5
Attach the excitation
plate of the filter cube
just mounted into the
rearmost pocket.
Filter cube switchover lever
Filter cube fixing plate
Attachment tool
Insert the excitation
plate to this pocket.