Chapter 7 Specifications
7.2 Microscope (Ti-E or Ti-E/B) with TI-DS Dia Pillar Illuminator 30W
125
7.2
Microscope (Ti-E or Ti-E/B) with TI-DS Dia Pillar Illuminator 30W
System configuration
For countries with power supply of 100 to 120 VAC:
Microscope, TI-DS Dia Pillar Illuminator 30W,
and TE-PS30W Power Supply A
For countries with power supply of 220 to 240 VAC:
Microscope, TI-DS Dia Pillar Illuminator 30W,
and TE-PSE30 Power Supply A
Dimensions
260 (W) x 497 (D) x 619 (H) mm
Mass
22.5 kg
Optical system
Objectives: CFI60
Eyepieces:
Field number 22
Nosepiece: Six
sockets
Mechanical system
Focusing mechanism:
Stroke: 10 mm
Focus knobs:
Coarse focus: 300
μ
m per rotation (reference value)*
Fine focus: 50
μ
m per rotation
Extra fine focus: 6.25
μ
m per rotation
*
When set to COARSE mode, the distance traveled by
rotating a focus knob will vary depending on how fast
the knob is rotated.
TI-DS
Dia Pillar Illuminator 30W
Input ratings:
6 VDC, 30 W
Lamp ratings:
6 V 30 W halogen lamp
Specified lamp model:
Halogen lamp (PHILIPS 5761)
Average lamp life:
100 hours
TE-PS30
Power Supply A
TE-PSE30
Power Supply A
Input ratings:
TE-PS30W: 100 to 120 VAC (±10%), 50/60 Hz, 0.7 A
TE-PSE30: 230 VAC (±10%), 50/60 Hz, 0.3 A
Built-in fuse ratings:
250 V F2AH
Output ratings:
6 VDC, 30 W
Maximum output current: 5.0 A
Electric shock protection class:
Class
I
Remarks:
TE-PS30W: UL listed product
TE-PSE30:
GS
approved
Operating conditions
Temperature:
0 to 40°C
Relative humidity:
85% RH max. with no condensation
Altitude:
2000 m max.
Pollution degree:
Degree 2
Installation category:
Category II
Indoor use only
Transport and storage
conditions
Temperature:
-20 to +60°C
Relative humidity:
90% RH max. with no condensation