(2) For TI-DS Dia Pillar Illuminator 30W
OPEN
TI-DS
30W
6V
30W
HALOGEN
Figure 4-33 Rear cover of the lamphouse
Figure 4-34 Replacing lamps
CAUTION: HOT
- Do not touch the lamp and
the lamphouse while the lamp is on or for
thirty minutes after it has been turned off.
1. Loosen the locking screw on the back of the
dia pillar illuminator, and remove the limit
plate from the rear cover.
2. Open the rear cover on the back of the dia
pillar illuminator by lifting it backward.
3. Remove the used lamp from the socket.
4. Insert a new lamp into the socket.
•
Be sure to use the specified lamp.
•
Do not touch the lamp glass with your bare
hands.
5. Close the rear cover.
6. Re-attach the limit plate onto the rear cover,
and secure it by tightening the locking screw.
Rear cover
Limit plate
Locking screw
Rear cover
Lamp