Chapter 4 Assembly
97
(3) Attaching the protection plate (when using an FL turret)
When using an FL turret with Ti-E or Ti-E/B, a protection plate must be attached between the FL turret and
the nosepiece. The required protection plate will differ depending on the nosepiece. Follow the procedures
below to attach the correct protection plate.
When using the TI-ND6-PFS PFS Motorized Nosepiece
The Protection Plate (for the PFS6 Nosepiece) supplied with the TI-ND6-PFS Perfect Focus Unit covers the
gap between the PFS Motorized Nosepiece and the FL turret (TI-FLC, TI-FLC-E, or TI-FLC-E/HQ), so as to
prevent injury caused by your hands and fingers getting caught.
When using an FL turret, attach the Protection Plate (for the PFS6 Nosepiece) before attaching the PFS
Motorized Nosepiece, as described below.
Figure 4-16 Protection Plate (for the PFS6
Nosepiece) attachment
1. Using a 2 mm hex screwdriver provided with
Ti-E or Ti-E/B, remove the four M3 hex socket
head screws from the top of the FL turret, as
shown in the figure at left.
2. Attach the Protection Plate (for the PFS6
Nosepiece) onto the top of the FL turret.
Insert and tighten the four M3 hex socket
head screws that were removed in step 1.
3. Pass the connection cable from the PFS
Motorized Nosepiece through the cable
through-hole on the protection plate. Using
the 4 mm hex screwdriver provided with Ti-E
or Ti-E/B, secure the nosepiece to the
microscope.
For details on attaching the PFS Motorized
Nosepiece, refer to “(1) Attaching the PFS
Motorized Nosepiece.”
When using other motorized nosepieces, refer to
“Motorized Units for Ti Series Instructions.”
Hexagonal socket
head screws and
washers (M5, x2)
PFS Motorized Nosepiece
FL turret
Protection Plate
(for the PFS6
Nosepiece)
Hexagonal socket
head screws
(M3, x4)